Eyelink 1000 calibration

From TSG Doc
Revision as of 13:30, 19 February 2016 by Wiki-admin (talk | contribs) (Created page with "==Calibration Eyelink 1000== <br /> ===Matlab=== Required hardware <br /> * Eyelink 1000 PC + camera * Extra computer with ethernet port (here referred to as PC2) Connection...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

Calibration Eyelink 1000


Matlab

Required hardware

  • Eyelink 1000 PC + camera
  • Extra computer with ethernet port (here referred to as PC2)

Connections

  • PC2 with ethernet cable to Eyelink PC (Note: use correct RJ45 input port)
  • Camera to Eyelink PC (big black cable)
  • Power supplies for computer and camera
  • Eyes left and right cables for camera should normally be and stay connected
  • Settings for ethernet connections: configure IP4 manually: ip-address: 100.1.1.2 and subnet mask: 255.255.255.0

Software

  • Required Eyelink framework library files: ask TSG to install these files.
  • Matlab and Psychtoolbox
  • BrainStream toolbox (also for for streaming Eyelink data to FieldTrip buffer, git@gitlab.socsci.ru.nl:cai-bci/brainstream.git)
  • Calibration function, available via BrainStream toolbox, $brainstream/resources/devices/eyelink/calibrate_eyelink.m

the calibration function includes an option to send an image to Eyelink PC monitor remotely via Eyelink connection and initialises stimulation screen fullscreen, which guarantees correct screen resolution parameters.
Note: start_eyelink_buffer script only after finishing the calibration procedure otherwise you run into a connection error


Instructions

  • start Matlab
  • change directory to $brainstream/resources/devices/eyelink ($brainstream refers to BrainStream toolbox folder)
  • add Psychtoolbox to Matlab path (you need to know the folder where Psychtoolbox is installed)
  • start:
  calibrate_eyelink(<optionally provide name of image to be shown as input argument>)
  • Calibration procedure starts/continues after pressing the ‘c’ key or the ‘calibrate' button on the Eyelink PC